Text Box: Technical Specification (PolAAr 31 and 3001)

Optical Activity Limited, Bury Road Industrial Estate, Ramsey, Cambridgeshire, PE26 1NF, United Kingdom
Tel: +44 (0)1487 813913   Fax: +44 (0)1487 812789
E-mail:  sales@opticalactivity.com   Web:  www.opticalactivity.com

Specification PolAAr 31 PolAAr 3001
Scales Angular degrees, ISS (°Z), temperature corrected ISS, specific rotation, concentration, 10 user programmable scales, apparent sugar purity if connected to an Index Instruments refractometer Angular degrees, ISS (°Z), temperature corrected ISS, specific rotation, concentration, 10 user programmable scales, apparent sugar purity if connected to an Index Instruments refractometer
Range Full circle, reading to ±90 angular degrees, in excess of ±250°Z Full circle, reading to ±90 angular degrees, in excess of ±250°Z
Resolution 0.01°,  0.01°Z 0.001°,  0.01°Z
Accuracy ±0.01°,  ±0.02°Z ±0.001° (0 to 10°)
±0.01° (10 to 90°)
±0.01°Z (0 to 25°Z)
±0.02°Z (above 25°Z)
Wavelengths 589.44nm¹ 589.44nm¹
Light Source 20 watt tungsten halogen lamp, typical life 2000 hours 20 watt tungsten halogen lamp, typical life 2000 hours
OD Tolerance OD 3.0 (ie 99.9% absorbance), warning issued if sample too dark OD 3.0 (ie 99.9% absorbance), warning issued if sample too dark
Reading Time (continuous mode) 10 seconds Approximately 10 to 15 seconds
Reading Modes Continuous, autoprint, one shot and lock, or read when stable Continuous, autoprint, one shot and lock, or read when stable
Sample Compartment Accepts standard sample tubes, 30mm diameter collars.  Maximum path length 200mm, bores 8mm down to 1.5mm.  Ventilated - temperature rise above ambient does not exceed 3°C Accepts standard sample tubes, 30mm diameter collars.  Maximum path length 200mm, bores 8mm down to 1.5mm.  Ventilated - temperature rise above ambient does not exceed 3°C
Thermometers  Temperature sensor in sample compartment, socket for alternative sample tube sensor.  Sensor range  0-100°C, accuracy ±0.25°C.  Digital display resolution 0.1°C Temperature sensor in sample compartment, socket for alternative sample tube sensor.  Sensor range  0-100°C, accuracy ±0.25°C.  Digital display resolution 0.1°C
Outputs/inputs 2 x RS232 (25-way D sockets), 1 x remote    (9-way D socket) 2 x RS232 (25-way D sockets), 1 x remote   (9-way D socket)
Data Output Rotation reading, units, time and date,  polarimeter serial number, wavelength, temperature, pre-set parameters, sample number Rotation reading, units, time and date,  polarimeter serial number, wavelength, temperature, pre-set parameters, sample number
Features 21 CFR Part ll Compliant 21 CFR Part ll Compliant
Special Features User selection and definition of output parameters for GLP (Good Laboratory Practice); in addition to the standard data output, up to 6 user defined sample identifiers     

User selection and definition of output parameters for GLP (Good Laboratory Practice); in addition to the standard data output, up to 6 user defined sample identifiers             

Controls Power ON/OFF, alpha-numeric keypad, plus function keys including print, zero and scale selection Power ON/OFF, alpha-numeric keypad, plus function keys including print, zero and scale selection
Calibration UKAS calibration certificate available if required UKAS calibration certificate available if required
Size and Weight L  515mm, D 380mm, H 198mm, W 15.7kg L  515mm, D 380mm, H 198mm, W 15.7kg

¹ Other wavelength options available